Products ElectrochemicalElectrochemical Workstation

Squidstat Venta

Make : Admiral Instruments, USA
Model : Squidstat Venta
Categories : Electrochemical,Electrochemical Workstation


 

±20 A Potentiostat with EIS up to 1 MHz

 

Squidstat Venta Highlights:

  • Becomes a multichannel system by connecting more Squidstats to the same computer
  • Onboard memory (16 GB) backs up all data collected from experiments for failsafe recovery
  • Voltage/current setpoints, OCP mode, and sampling rates are adjustable in real time
  • Wide selection of prebuilt experiments are available, and the “Drag & Drop” experiment builder can quickly and easily make custom experiments – no coding skills necessary!

  • Description
  • Specifications


 

±20 A Potentiostat with EIS up to 1 MHz

 

Squidstat Venta Highlights:

  • Becomes a multichannel system by connecting more Squidstats to the same computer
  • Onboard memory (16 GB) backs up all data collected from experiments for failsafe recovery
  • Voltage/current setpoints, OCP mode, and sampling rates are adjustable in real time
  • Wide selection of prebuilt experiments are available, and the “Drag & Drop” experiment builder can quickly and easily make custom experiments – no coding skills necessary!

General Specifications

 

 

Channels Per Unit

1

Operating Modes

Pot / Gal / ZRA / FRA

Cell Connections

2, 3, 4, or 5 electrode

ADC/DAC Resolution

16 bit resolution

Slew Rate (No Load)

5 MV/s

Maximum Sampling Speed

100,000 samples/s

Floating and/or Earth-grounding

Both (Switchable)

Input Impedance

> 10 Tera-ohm

Input Bias Current

< 1 pA

Data Backup Memory

16 GB

Physical Dimensions

23 cm x 17 cm x 13 cm

Channel Cable Length

100 cm

Computer Interface

1 USB per unit

Power Supply Requirements

100 – 240 VAC

50 – 60 Hz

 

Potentiostatic Input/Output

Compliance Voltage

±6 V

Voltage Scan Range

±6 V

Number of Voltage Ranges

3 ranges,

(0.12 V, 1.2V, 6 V)

Applied Potential Accuracy

0.1% of setpoint,

1 mV min

Applied Potential Resolution

0.003% of range,

3.6 μV min

Measured Potential Accuracy

0.1% of reading,

1 mV min

Measured Potential Resolution

0.003% of range,

3.6 μV min

 

Galvanostatic Input/Output

Maximum Current

±20 A

Current Ranges

9 ranges,

(1 μA to 20 A)

Applied Current Accuracy

0.1% of range,

1 nA min

Applied Current Resolution

0.003% of range,

30 pA min

Measured Current Accuracy

0.1% of range,

1 nA min

Measured Current Resolution

0.003% of range,

30 pA min

 

Impedance Analyzer

EIS Frequency Range

10 μHz to 1 MHz

AC Frequency Accuracy

0.005% or better

AC Frequency Resolution

0.0004%, 3 µHz min

Potentiostatic Amplitude

≤ 100% of range,

6 V max

Potentiostatic

Amplitude

Resolution

25 µV @ 1mV

250 µV @ 100mV

1.94 mV @ 6V

Galvanostatic Amplitude

≤ 100% of range,

20 A max

Galvanostatic

Amplitude

Resolution

0.025% of applied @ 100% of range
0.08% of applied @ 10% of range

Career