Model: |
BTS- |
5V20A |
Model: |
BTS- |
5V20A |
Model: |
BTS- |
5V20A |
Description |
Product Specification |
||
AC Input |
AC 220V±10% 50Hz |
||
Power |
1418 |
W |
|
Resolution |
AD: 16bit; DA: 16bit |
||
Input Impedance |
≥1MΩ |
||
Voltage |
Measuring Range |
25mV~5V |
|
Discharge Min Voltage |
2.5V |
||
Accuracy |
± 0.1% of FS |
||
Stability |
± 0.1% of FS |
||
Current |
Range |
Charge:100mA~20A Discharge:100mA~20A |
|
Accuracy |
± 0.1% of FS |
||
Stability |
± 0.1% of FS |
||
Power |
Output Power Per Channel |
100W |
|
Stability |
± 0.2% of FS |
||
Aux Chanel |
Aux Input |
Temperature |
|
Temperature |
-25℃~110℃ |
||
Temperature Accuracy |
±2℃ |
||
Temperature Resolution |
0.1℃ |
||
Time |
Rise Time |
20ms (10%~90%FS) |
|
Step Time |
Time Format 00:00:00:000 ≤ (365*24) hour/step (h:min:s: ms) |
||
Data Acquisition |
Intervals |
Time interval Δt:(100ms~60000s) |
|
Voltage interval ΔU: (10mV~5V) |
|||
Current interval ΔI: 40mA~20A |
|||
Frequency |
10Hz |
||
Charge |
Mode of Operation |
CCC, CVC, CC & CVC, CPC |
|
End Conditions |
Voltage, Current, Test Time, Capacity |
||
Discharge |
Mode of Operation |
CCD, CPD, CRD |
|
End Conditions |
Voltage, Current, Test Time, Capacity |
Pulse mode |
Charge |
CCC |
Discharge |
CCD |
|
Min pulse width |
500ms |
|
Automated Switch |
Automated switch from charge to discharge for each pulse |
|
End Condition |
Voltageã€Test Timeã€Currentã€Capacity |
|
DCIR |
Support for the calculation of the point of the DCIR |
|
Cycle |
Cycles |
65535 |
Steps |
254 |
|
Nested Function |
Max three levels of loops |
|
Protective Function |
Software |
|
|
||
voltage and delay time |
||
Channel Features |
Independent pairs of a closed loop for the constant current source and constant voltage source |
|
Channels |
Independent control |
|
Detection and Sampling |
4-wire Connecting |
|
Noise Density |
<85dB |
|
Data Management |
MYSQL Database |
|
Communication Means |
TCP/IP Protocol |
|
Export Formats |
EXCEL, TXT, Graph |
|
Communication Interface |
Ethernet Port |
|
Number of Channels Per Cabinet |
8 |
|
Operating Environment |
||
Description |
Product Specification |
|
Operating Temperature |
0℃~40℃ |
|
Storage Temperature |
-10℃~50℃ |
|
Operating Humidity |
≤70% RH |
|
Storage Humidity |
≤80% RH |
|
Description |
Product Specification |
|
Battery Holder |
Universal Holder |